Two different hematocrit detection methods: Different methods, different results?


Bosshart, M; Stover, J F; Stocker, R; Asmis, L M; Feige, J; Neff, T A; Schuepbach, R A; Cottini, S R; Béchir, M (2010). Two different hematocrit detection methods: Different methods, different results? BMC Research Notes, 3:65.

Abstract

BACKGROUND: Less is known about the influence of hematocrit detection methodology on transfusion triggers. Therefore, the aim of the present study was to compare two different hematocrit-assessing methods. In a total of 50 critically ill patients hematocrit was analyzed using (1) blood gas analyzer (ABLflex 800) and (2) the central laboratory method (ADVIA(R) 2120) and compared. FINDINGS: Bland-Altman analysis for repeated measurements showed a good correlation with a bias of +1.39% and 2 SD of +/- 3.12%. The 24%-hematocrit-group showed a correlation of r2 = 0.87. With a kappa of 0.56, 22.7% of the cases would have been transfused differently. In the-28%-hematocrit group with a similar correlation (r2 = 0.8) and a kappa of 0.58, 21% of the cases would have been transfused differently. CONCLUSIONS: Despite a good agreement between the two methods used to determine hematocrit in clinical routine, the calculated difference of 1.4% might substantially influence transfusion triggers depending on the employed method.

Abstract

BACKGROUND: Less is known about the influence of hematocrit detection methodology on transfusion triggers. Therefore, the aim of the present study was to compare two different hematocrit-assessing methods. In a total of 50 critically ill patients hematocrit was analyzed using (1) blood gas analyzer (ABLflex 800) and (2) the central laboratory method (ADVIA(R) 2120) and compared. FINDINGS: Bland-Altman analysis for repeated measurements showed a good correlation with a bias of +1.39% and 2 SD of +/- 3.12%. The 24%-hematocrit-group showed a correlation of r2 = 0.87. With a kappa of 0.56, 22.7% of the cases would have been transfused differently. In the-28%-hematocrit group with a similar correlation (r2 = 0.8) and a kappa of 0.58, 21% of the cases would have been transfused differently. CONCLUSIONS: Despite a good agreement between the two methods used to determine hematocrit in clinical routine, the calculated difference of 1.4% might substantially influence transfusion triggers depending on the employed method.

Statistics

Citations

Dimensions.ai Metrics

Altmetrics

Downloads

118 downloads since deposited on 15 Nov 2010
10 downloads since 12 months
Detailed statistics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Institute of Intensive Care Medicine
04 Faculty of Medicine > University Hospital Zurich > Clinic for Oncology and Hematology
04 Faculty of Medicine > University Hospital Zurich > Institute of Anesthesiology
Dewey Decimal Classification:610 Medicine & health
Scopus Subject Areas:Life Sciences > General Biochemistry, Genetics and Molecular Biology
Language:English
Date:2010
Deposited On:15 Nov 2010 13:29
Last Modified:28 Jun 2022 10:50
Publisher:BioMed Central
ISSN:1756-0500
OA Status:Gold
Free access at:PubMed ID. An embargo period may apply.
Publisher DOI:https://doi.org/10.1186/1756-0500-3-65
PubMed ID:20214819

Download

Gold Open Access

Download PDF  'Two different hematocrit detection methods: Different methods, different results?'.
Preview
Content: Published Version
Filetype: PDF
Size: 283kB
View at publisher
Licence: Creative Commons: Attribution 2.0 Generic (CC BY 2.0)