Evaluation of the Mythic 18, haematology analyser for its use in dogs, cats and horses


Wassmuth, Andrea Katharina. Evaluation of the Mythic 18, haematology analyser for its use in dogs, cats and horses. 2010, University of Zurich, Vetsuisse Faculty.

Abstract

The Mythic 18 is a fully automated haematology bench-top analyser using impedance technology for a complete blood count (CBC) and a 3-part white blood cell (WBC) differential. The purpose of this study was to evaluate the Mythic for accuracy, precision, linearity, carry-over, stability and usability under practice conditions. EDTA- blood samples from 122 dogs, 140 cats and 123 horses were analysed with the Mythic and reference methods (Sysmex XT-2000iV, manual haematocrit and microscopic WBC differentiation). Red blood cell parameters showed excellent correlation and small biases. Total WBC count correlated excellently in canine and equine and very well in feline samples. In 23 feline specimens with platelet aggregates, the Mythic overestimated WBC counts. In all three species, absolute granulocyte counts correlated excellently. Equine lymphocyte counts showed good correlation whereas canine and feline lymphocyte counts correlated poorly. Feline platelets showed good correlation with a negative bias. The instrument showed good to excellent precision and performed excellently for the CBC count parameters in all investigated species. The whole 3-part differential was found to be accurate in horses. In dogs and cats absolute granulocyte counts were reliable. As with all impedance based haematological instruments, evaluation of a blood smear is absolutely indicated to check for the presence of platelet aggregates, to verify WBC differentiation and to identify possible pathologies.

Evaluierung des Hämatologiegerätes Mythic 18 für Katzen, Hunde und Pferde

Der Mythic 18 ist ein vollautomatisches Hämatologiegerät für den Gebrauch in der tierärztlichen Praxis. Basierend auf der Impedanztechnik analysiert das Gerät 16 Parameter inklusive einer 3-Part Leukozytendifferenzierung (Lymphozyten, Monozyten und Granulozyten). Das Ziel dieser Studie war es, für den Mythic Richtigkeit, Präzision, Linearität, den „Carry-over“-Wert und die Stabilität zu bestimmen. Blutproben von 122 Hunden, 140 Katzen und 123 Pferden wurden im Rahmen einer vergleichenden Untersuchung mit dem Mythic und Referenzmethoden (Sysmex XT-2000iV, manueller Hämatokrit und manueller WBC Differenzierung) analysiert. Die Parameter des roten Blutbilds wurden durch den Mythic richtig und präzise bestimmt. Die absoluten Leukozytenwerte korrelierten exzellent bei Hunde- und Pferdeproben. Bei der Katze zeigte sich eine geringe Abweichung. Der Mythic 18 überschätzte allerdings die Leukozytenzahlen bei 23 Proben mit Thrombozytenaggregaten. Bei allen drei Tierarten korrelierten die absoluten Granulozytenwerte sehr gut mit den Referenzmethoden. Die Präzision des Gerätes war sehr gut. Der Mythic zeigte bei allen untersuchten Tierarten für das kleine Blutbild eine sehr gute Übereinstimmung mit jenen des Referenzgerätes. Das Differenzialblutbild erwies sich als gut brauchbar für Pferde. Die Evaluierung eines Blutausstriches in Bezug auf das Vorhandensein von Thrombozytenaggregaten, der Verifizierung der Leukozytendifferenzierung und zur Erkennung von Pathologien ist notwendig.

Abstract

The Mythic 18 is a fully automated haematology bench-top analyser using impedance technology for a complete blood count (CBC) and a 3-part white blood cell (WBC) differential. The purpose of this study was to evaluate the Mythic for accuracy, precision, linearity, carry-over, stability and usability under practice conditions. EDTA- blood samples from 122 dogs, 140 cats and 123 horses were analysed with the Mythic and reference methods (Sysmex XT-2000iV, manual haematocrit and microscopic WBC differentiation). Red blood cell parameters showed excellent correlation and small biases. Total WBC count correlated excellently in canine and equine and very well in feline samples. In 23 feline specimens with platelet aggregates, the Mythic overestimated WBC counts. In all three species, absolute granulocyte counts correlated excellently. Equine lymphocyte counts showed good correlation whereas canine and feline lymphocyte counts correlated poorly. Feline platelets showed good correlation with a negative bias. The instrument showed good to excellent precision and performed excellently for the CBC count parameters in all investigated species. The whole 3-part differential was found to be accurate in horses. In dogs and cats absolute granulocyte counts were reliable. As with all impedance based haematological instruments, evaluation of a blood smear is absolutely indicated to check for the presence of platelet aggregates, to verify WBC differentiation and to identify possible pathologies.

Evaluierung des Hämatologiegerätes Mythic 18 für Katzen, Hunde und Pferde

Der Mythic 18 ist ein vollautomatisches Hämatologiegerät für den Gebrauch in der tierärztlichen Praxis. Basierend auf der Impedanztechnik analysiert das Gerät 16 Parameter inklusive einer 3-Part Leukozytendifferenzierung (Lymphozyten, Monozyten und Granulozyten). Das Ziel dieser Studie war es, für den Mythic Richtigkeit, Präzision, Linearität, den „Carry-over“-Wert und die Stabilität zu bestimmen. Blutproben von 122 Hunden, 140 Katzen und 123 Pferden wurden im Rahmen einer vergleichenden Untersuchung mit dem Mythic und Referenzmethoden (Sysmex XT-2000iV, manueller Hämatokrit und manueller WBC Differenzierung) analysiert. Die Parameter des roten Blutbilds wurden durch den Mythic richtig und präzise bestimmt. Die absoluten Leukozytenwerte korrelierten exzellent bei Hunde- und Pferdeproben. Bei der Katze zeigte sich eine geringe Abweichung. Der Mythic 18 überschätzte allerdings die Leukozytenzahlen bei 23 Proben mit Thrombozytenaggregaten. Bei allen drei Tierarten korrelierten die absoluten Granulozytenwerte sehr gut mit den Referenzmethoden. Die Präzision des Gerätes war sehr gut. Der Mythic zeigte bei allen untersuchten Tierarten für das kleine Blutbild eine sehr gute Übereinstimmung mit jenen des Referenzgerätes. Das Differenzialblutbild erwies sich als gut brauchbar für Pferde. Die Evaluierung eines Blutausstriches in Bezug auf das Vorhandensein von Thrombozytenaggregaten, der Verifizierung der Leukozytendifferenzierung und zur Erkennung von Pathologien ist notwendig.

Statistics

Downloads

1291 downloads since deposited on 10 Feb 2011
105 downloads since 12 months
Detailed statistics

Additional indexing

Item Type:Dissertation (monographical)
Referees:Lutz Hans, Lutz Thomas A
Communities & Collections:05 Vetsuisse Faculty > Veterinary Clinic > Department of Farm Animals
UZH Dissertations
Dewey Decimal Classification:570 Life sciences; biology
630 Agriculture
Language:English
Place of Publication:Zürich
Date:2010
Deposited On:10 Feb 2011 11:27
Last Modified:24 Sep 2019 17:23
Number of Pages:43
Additional Information:Evaluation of the Mythic 18, haematology analyser for its use in dogs, cats and horses / vorgelegt von Andrea Katharina Wassmuth. - Zürich, 2010
OA Status:Green
Related URLs:https://www.recherche-portal.ch/primo-explore/fulldisplay?docid=ebi01_prod006170033&context=L&vid=ZAD&search_scope=default_scope&tab=default_tab&lang=de_DE (Library Catalogue)

Download

Green Open Access

Download PDF  'Evaluation of the Mythic 18, haematology analyser for its use in dogs, cats and horses'.
Preview
Filetype: PDF
Size: 5MB