Use of a single bipolar electrode in the posterior arytenoid muscles for bilateral monitoring of the recurrent laryngeal nerves in thyroid surgery


Haerle, S; Sidler, D; Linder, T; Mueller, W (2008). Use of a single bipolar electrode in the posterior arytenoid muscles for bilateral monitoring of the recurrent laryngeal nerves in thyroid surgery. European Archives of Oto-Rhino-Laryngology, 265(12):1549-1552.

Abstract

The aims were to assess the technical feasibility of using a single electrode in the posterior arytenoid muscles (PAM) for intraoperative monitoring of the recurrent laryngeal nerve (RLN) in thyroid surgery, to validate the new method against the insertion of electrodes placed in the vocal cord muscle, and to report the results of the clinical application of the new concept. A total of 52 patients were enrolled. The handling and safety of RLN monitoring was tested by simultaneous registration of the EMG response from vocal fold electrodes and PAM electrodes. Acoustically and electromyographically we found nearly the same values for the arytenoid muscles as for the vocal folds, although the signals taken from the vocal folds were slightly stronger. PAM recording using a single bipolar electrode is technically feasible and as reliable compared to the standard vocal cord monitoring.

Abstract

The aims were to assess the technical feasibility of using a single electrode in the posterior arytenoid muscles (PAM) for intraoperative monitoring of the recurrent laryngeal nerve (RLN) in thyroid surgery, to validate the new method against the insertion of electrodes placed in the vocal cord muscle, and to report the results of the clinical application of the new concept. A total of 52 patients were enrolled. The handling and safety of RLN monitoring was tested by simultaneous registration of the EMG response from vocal fold electrodes and PAM electrodes. Acoustically and electromyographically we found nearly the same values for the arytenoid muscles as for the vocal folds, although the signals taken from the vocal folds were slightly stronger. PAM recording using a single bipolar electrode is technically feasible and as reliable compared to the standard vocal cord monitoring.

Statistics

Citations

Dimensions.ai Metrics
5 citations in Web of Science®
3 citations in Scopus®
Google Scholar™

Altmetrics

Downloads

65 downloads since deposited on 04 Nov 2008
9 downloads since 12 months
Detailed statistics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Clinic for Otorhinolaryngology
Dewey Decimal Classification:610 Medicine & health
Scopus Subject Areas:Health Sciences > Otorhinolaryngology
Language:English
Date:2008
Deposited On:04 Nov 2008 16:52
Last Modified:01 Dec 2023 02:46
Publisher:Springer
ISSN:0937-4477
OA Status:Green
Publisher DOI:https://doi.org/10.1007/s00405-008-0671-3
PubMed ID:18415115
  • Content: Published Version
  • Language: English
  • Description: Nationallizenz 142-005