Low-dose multiple-information retrieval algorithm for X-ray grating-based imaging


Wang, Zhentian; Huang, Zhifeng; Chen, Zhiqiang; Zhang, Li I; Jiang, Xiaolei; Kang, Kejun; Yin, Hongxia; Wang, Zhenchang; Stampanoni, Marco (2011). Low-dose multiple-information retrieval algorithm for X-ray grating-based imaging. Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 635(1):103-107.

Abstract

The present work proposes a low dose information retrieval algorithm for X-ray grating-based multiple-information imaging (GB-MII) method, which can retrieve the attenuation, refraction and scattering information of samples by only three images. This algorithm aims at reducing the exposure time and the doses delivered to the sample. The multiple-information retrieval problem in GB-MII is solved by transforming a nonlinear equations set to a linear equations and adopting the nature of the trigonometric functions. The proposed algorithm is validated by experiments both on conventional X-ray source and synchrotron X-ray source, and compared with the traditional multiple-image-based retrieval algorithm. The experimental results show that our algorithm is comparable with the traditional retrieval algorithm and especially suitable for high Signal-to-Noise system.

Abstract

The present work proposes a low dose information retrieval algorithm for X-ray grating-based multiple-information imaging (GB-MII) method, which can retrieve the attenuation, refraction and scattering information of samples by only three images. This algorithm aims at reducing the exposure time and the doses delivered to the sample. The multiple-information retrieval problem in GB-MII is solved by transforming a nonlinear equations set to a linear equations and adopting the nature of the trigonometric functions. The proposed algorithm is validated by experiments both on conventional X-ray source and synchrotron X-ray source, and compared with the traditional multiple-image-based retrieval algorithm. The experimental results show that our algorithm is comparable with the traditional retrieval algorithm and especially suitable for high Signal-to-Noise system.

Statistics

Citations

Dimensions.ai Metrics
20 citations in Web of Science®
24 citations in Scopus®
Google Scholar™

Altmetrics

Downloads

0 downloads since deposited on 01 Mar 2012
0 downloads since 12 months

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > Institute of Biomedical Engineering
Dewey Decimal Classification:170 Ethics
610 Medicine & health
Scopus Subject Areas:Physical Sciences > Nuclear and High Energy Physics
Physical Sciences > Instrumentation
Language:English
Date:2011
Deposited On:01 Mar 2012 14:25
Last Modified:07 Nov 2023 02:43
Publisher:Elsevier
ISSN:0168-9002
OA Status:Closed
Publisher DOI:https://doi.org/10.1016/j.nima.2011.01.079