Intraoperative assessment of stapes movement


Huber, A; Linder, T; Ferrazzini, M; Schmid, S; Dillier, N; Stoeckli, S; Fisch, U (2001). Intraoperative assessment of stapes movement. Annals of Otology, Rhinology and Laryngology, 110(1):31-35.

Abstract

A method is described that allows, for the first time, intraoperative vibration modes assessment of the acoustically stimulated stapes by means of scanning laser Doppler interferometry (LDI). The study was designed to answer the following questions: 1) Is LDI practical for taking measurements during surgery? 2) Are the results comparable to the findings in temporal bone preparations? and 3) Do the vibration characteristics of the stapes change after the posterior incudal ligament is detached from the incus? Seven patients with profound bilateral hearing loss who were undergoing cochlear implantation were included in the study. The measurement system was easily applicable for intraoperative measurements and allowed contact-free analysis with very high accuracy. No major differences in the results from the live human subjects and temporal bone preparations were observed. The stapes movement was predominantly pistonlike at the lower frequencies and became complex at higher frequencies. Sacrificing the posterior incudal ligament had no statistically significant effect on stapes vibration.

Abstract

A method is described that allows, for the first time, intraoperative vibration modes assessment of the acoustically stimulated stapes by means of scanning laser Doppler interferometry (LDI). The study was designed to answer the following questions: 1) Is LDI practical for taking measurements during surgery? 2) Are the results comparable to the findings in temporal bone preparations? and 3) Do the vibration characteristics of the stapes change after the posterior incudal ligament is detached from the incus? Seven patients with profound bilateral hearing loss who were undergoing cochlear implantation were included in the study. The measurement system was easily applicable for intraoperative measurements and allowed contact-free analysis with very high accuracy. No major differences in the results from the live human subjects and temporal bone preparations were observed. The stapes movement was predominantly pistonlike at the lower frequencies and became complex at higher frequencies. Sacrificing the posterior incudal ligament had no statistically significant effect on stapes vibration.

Statistics

Citations

Dimensions.ai Metrics
85 citations in Web of Science®
83 citations in Scopus®
Google Scholar™

Altmetrics

Downloads

146 downloads since deposited on 30 Mar 2009
23 downloads since 12 months
Detailed statistics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Clinic for Otorhinolaryngology
Dewey Decimal Classification:610 Medicine & health
Scopus Subject Areas:Health Sciences > Otorhinolaryngology
Language:English
Date:2001
Deposited On:30 Mar 2009 11:44
Last Modified:23 Jan 2022 12:47
Publisher:Annals Publishing
ISSN:0003-4894
OA Status:Green
Publisher DOI:https://doi.org/10.1177/000348940111000106
PubMed ID:11201805