Mechanical versus manual chest compression CPR under ground ambulance transport conditions


Fox, Julia; Fiechter, René; Gerstl, Peter; Url, Alfons; Wagner, Heinz; Lüscher, Thomas F; Eriksson, Urs; Wyss, Christophe A (2013). Mechanical versus manual chest compression CPR under ground ambulance transport conditions. Acute Cardiac Care, 15(1):1-6.

Abstract

BACKGROUND: Novel mechanical chest compression devices offer the possibility to transport cardiac arrest patients with ongoing CPR and might shorten significantly the time delay to post-resuscitation care.
METHODS: We simulated an eight-minute cardiac resuscitation situation during ambulance transport using CPR training manikins. We compared teams consisting of two experienced resuscitators with the performance of a mechanical chest compression device (LUCAS).
RESULTS: CPR-performance by two experienced resuscitators demonstrated ambivalent results. Whereas mean compression rate was within the recommended range (103/min, 95% CI: 93-113/min), mean compression depth was closely below the actually recommended compression depth of >5 cm (49.7 mm, 95% CI: 46.1-53.3mm). Nevertheless, only a mean of two thirds (67%) of all compressions were classified as manually correct (defined as sternal compression depth >5 cm). In contrast, the LUCAS device showed a constant and reliable CPR performance (99.96% correctly applied chest compressions correctly applied within the device programmed parameters, P = 0.0162) with almost no variance between the different sequences.
CONCLUSION: The LUCAS CPR device represents a reliable alternative to manual CPR in a moving ambulance vehicle during emergency evacuation. Furthermore, it needs less human resources and is safer for the EMS personnel.

Abstract

BACKGROUND: Novel mechanical chest compression devices offer the possibility to transport cardiac arrest patients with ongoing CPR and might shorten significantly the time delay to post-resuscitation care.
METHODS: We simulated an eight-minute cardiac resuscitation situation during ambulance transport using CPR training manikins. We compared teams consisting of two experienced resuscitators with the performance of a mechanical chest compression device (LUCAS).
RESULTS: CPR-performance by two experienced resuscitators demonstrated ambivalent results. Whereas mean compression rate was within the recommended range (103/min, 95% CI: 93-113/min), mean compression depth was closely below the actually recommended compression depth of >5 cm (49.7 mm, 95% CI: 46.1-53.3mm). Nevertheless, only a mean of two thirds (67%) of all compressions were classified as manually correct (defined as sternal compression depth >5 cm). In contrast, the LUCAS device showed a constant and reliable CPR performance (99.96% correctly applied chest compressions correctly applied within the device programmed parameters, P = 0.0162) with almost no variance between the different sequences.
CONCLUSION: The LUCAS CPR device represents a reliable alternative to manual CPR in a moving ambulance vehicle during emergency evacuation. Furthermore, it needs less human resources and is safer for the EMS personnel.

Statistics

Citations

Dimensions.ai Metrics

Altmetrics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Clinic for Cardiology
Dewey Decimal Classification:610 Medicine & health
Scopus Subject Areas:Health Sciences > Emergency Medicine
Health Sciences > Radiology, Nuclear Medicine and Imaging
Health Sciences > Cardiology and Cardiovascular Medicine
Language:English
Date:2013
Deposited On:12 Feb 2014 15:21
Last Modified:24 Jan 2022 03:20
Publisher:Informa Healthcare
ISSN:1748-2941
OA Status:Closed
Publisher DOI:https://doi.org/10.3109/17482941.2012.735675
PubMed ID:23425006
Full text not available from this repository.