Header

UZH-Logo

Maintenance Infos

Computed tomography for planning and postoperative imaging of transvenous mitral annuloplasty: first experience in an animal model


Sündermann, Simon H; Gordic, Sonja; Manka, Robert; Cesarovic, Nikola; Falk, Volkmar; Maisano, Francesco; Alkadhi, Hatem (2015). Computed tomography for planning and postoperative imaging of transvenous mitral annuloplasty: first experience in an animal model. International Journal of Cardiovascular Imaging, 31(1):135-142.

Abstract

To investigate the use of computed tomography (CT) to measure the mitral valve annulus size before implantation of a percutaneous mitral valve annuloplasty device in an animal trial. Seven domestic pigs underwent CT before and after implantation of a Cardioband™ (a percutaneously implantable mitral valve annuloplasty device) with a second-generation 128-section dual-source CT machine. Implantation of the Cardioband™ was performed in a standard fashion according to a protocol. Animals were sacrificed afterwards and the hearts explanted. The Cardioband™ was found to be adequately implanted in all animals, with no anchor dehiscence and no damage of the circumflex artery (CX) or the coronary sinus (CS). The correct length of the band as chosen according to the length of the posterior mitral annulus measured in CT before implantation was confirmed in gross examination in all animals. The device did not result in a metal artifact-related degradation of image quality. The closest distance from the closest anchor to the CX was 2.1 ± 0.7 mm in diastole and 1.6 ± 0.5 mm systole. Mitral annulus distance to the CS was 6.4 ± 1.3 mm in diastole and 7.7 ± 1.1 mm in systole. CT visualization and measurement of the mitral valve annulus dimensions is feasible and can become the imaging method of choice for procedure planning of Cardioband™ implantations or other transcatheter mitral annuloplasty devices.

Abstract

To investigate the use of computed tomography (CT) to measure the mitral valve annulus size before implantation of a percutaneous mitral valve annuloplasty device in an animal trial. Seven domestic pigs underwent CT before and after implantation of a Cardioband™ (a percutaneously implantable mitral valve annuloplasty device) with a second-generation 128-section dual-source CT machine. Implantation of the Cardioband™ was performed in a standard fashion according to a protocol. Animals were sacrificed afterwards and the hearts explanted. The Cardioband™ was found to be adequately implanted in all animals, with no anchor dehiscence and no damage of the circumflex artery (CX) or the coronary sinus (CS). The correct length of the band as chosen according to the length of the posterior mitral annulus measured in CT before implantation was confirmed in gross examination in all animals. The device did not result in a metal artifact-related degradation of image quality. The closest distance from the closest anchor to the CX was 2.1 ± 0.7 mm in diastole and 1.6 ± 0.5 mm systole. Mitral annulus distance to the CS was 6.4 ± 1.3 mm in diastole and 7.7 ± 1.1 mm in systole. CT visualization and measurement of the mitral valve annulus dimensions is feasible and can become the imaging method of choice for procedure planning of Cardioband™ implantations or other transcatheter mitral annuloplasty devices.

Statistics

Citations

Dimensions.ai Metrics
15 citations in Web of Science®
14 citations in Scopus®
Google Scholar™

Altmetrics

Downloads

12 downloads since deposited on 19 Aug 2014
12 downloads since 12 months
Detailed statistics

Additional indexing

Item Type:Journal Article, refereed, original work
Communities & Collections:04 Faculty of Medicine > University Hospital Zurich > Clinic for Diagnostic and Interventional Radiology
04 Faculty of Medicine > Institute of Biomedical Engineering
Dewey Decimal Classification:170 Ethics
610 Medicine & health
Scopus Subject Areas:Health Sciences > Radiology, Nuclear Medicine and Imaging
Health Sciences > Cardiology and Cardiovascular Medicine
Language:English
Date:2015
Deposited On:19 Aug 2014 14:17
Last Modified:24 Jan 2022 04:35
Publisher:Springer
ISSN:1569-5794
OA Status:Green
Publisher DOI:https://doi.org/10.1007/s10554-014-0516-7
PubMed ID:25119889

Download

Green Open Access

Download PDF  'Computed tomography for planning and postoperative imaging of transvenous mitral annuloplasty: first experience in an animal model'.
Preview
Content: Published Version
Language: English
Filetype: PDF (Nationallizenz 142-005)
Size: 818kB
View at publisher